Instrument Cradle

Homepage GUEST BOOK - SIGN IN Contact Me Published Articles & Speaking Benchmarking  Clean Steam Custom Pack Class-6 Indicators Favorite Links Favorite Articles Flash Sterilization Forms For Sterilization Instrument Cradle Instrument Management Impingement Washing Island Sink Just-In-Time New Stuff OR Business Management Rigid Scopes Service Contract Storage Bins Suture Carts What's Wrong With CS? Who’s Managing CS? Who's Managing the OR? What's New Legal Notice

Improving The Overall Cleaning Process

For years, we in CSSPD have been trying to find a method to clean instruments with minimal contact of bio-burden to our employees.  We have also been instructed to open all ring-handled instruments and arrange them so that full water impingement can occur in the washer disinfector.

One attempt at this was the 6-inch stringer which meant that all ring-handled instruments had to be pre-strung on the dirty side and de-strung on the clean side to complete PM testing of scissors, needleholders, and clamps.  If used properly the 6-inch strimger did improve the washer process as long as the instruments were placed in an upright position, not laying on one side or the other. 

This added additional time to both the pre and post washer and was the main reason why most hospitals did not adapt the practice.  If you pre-string then you must have a well defined manual process starting in the OR with soaking and moving forward to cleaning following the stringer manufacturer’s recommendations for the angle of the stringed instruments, again preferably standing up-right to allow full impingement.

 

Full Washer Impingement

With impingement representing about 60% of the washers ability to remove contaminates the need to develope a product to enhanced the washers ability has never been more needed.  

The introduction of the instrument cradle answered all that. This simple device allows full water impingement in a washer disinfector as well as proper drainage and faster cooling.  The time spent pre-stringing is eliminated while improving  assembly time due to the instruments becoming more visible to the CS tech.

Contact to Target Surface

Your home dish washer is designed to place everything in an up-right position allowing for fully contact and drainage.  Not one of the current manufacturers of hospital washers have a racking system to solve what the home industry has, load configurations and limits.

Reduces the need for extensive hand washing of instruments

If you are experiencing spotting or staining on surgical instruments especially in the box-lock area of ring-handled instruments the reason is due to lack of water impingement in the washer disinfector and a poor manual process.  

A poor pre-washer, soak, and clean process adds to the problem.

Washer Impingement Improved When Using The Cradle

Impingement efficiency is dependent on contact to the primary surface area which is directly related to the placement and positioning of instruments.  Surgical instrumentations primary area is the surface that has the highest contact with human tissue including hinged and uni-body areas.  

High Impingement is more successful when direct contact can be made on the targeted surface.  Positioning of instruments becomes even more challenging when the quantity or mass of instruments exceeds the washer’s ability and there are no limits set. 

Improved Flow Enhances Drainage and Post-Washer Cooling

Furthermore, if you can see visible staining, rusting, or pitting then you have a high potential for forming bio-film and something needs to be done to correct it.   Visible rust on a surgical instruments means that your process is failing and that all levels of cleaning need to be addressed.

  • Sorting
  • Disassembly
  • Soaking
  • Poor manual hand washing
  • Improper racking
  • Plugged spray arms
  • Washer chemical level to low
  • DI/RO rinse water cycle to short
  • Over use of flash sterilization along with poor manual hand washing technique.
  • Using surgical towels in wraped or containerized instrument sets
  • Over use of rust remover chemicals
  • Overloading the washer racks
  • Poor rack selection
  • Poor placement on chosen rack
  • layering of instruments and use of hold-down screen   

No Evidence Based Studies To Support Current Instrument Cleaning Process

A breakdown in the passivation layer due to poor instrument management will result in staining and spotting.

Having your instrument repair service complete routine passivation as part of the monthly maintenance service along with a quality cleaning process will eliminate this.

Relying on automated washers would be your first mistake in correcting the problem.  The use of the cradle will greatly enhance your washer’s ability to properly disinfect ring-handled instruments which are the bulk of your instrument inventory.

Box-lock with staining and rusting that should never make it to the sterile field.

Laying stringed instruments down rather then in an up-right position completely eleminates the wahers ability to make contact with the target surface or tissue contact area, serrations and teeth.

We have been using the cradle at our hospital for eighteen months and have seen a reduction in repairs as well as remarkable increase in the look and feel of ring-handled instruments. 

No visible staining or spotting problems,

No complaints from staff regarding the use of the cradle, 

No employee sharps injuries for the past eighteen months while using the cradle both pre and post washer during tray assembly.

CLICK TO ENLARGE

For direct infromation contact:

www.trucleaninnovations.com

Call Michele Bell at Tru-Clean Innovations - 1-866-453-1842

http://csspdmanager.com/Cradle/TruClean_V1-2.wmv